I2C protocol

The SF06 command set as the driver uses it: framing, every command, the frames the sensor returns, scaling and the status word.

Source: SFM4300 datasheet, section 4. The constants are in inc/sfm_commands.hpp.

Framing

  • A command is 16 bits, MSB first, written without a register byte.
  • A command with an argument adds a 16-bit argument and its CRC-8: five bytes in all.
  • A read returns 16-bit words, each followed by a CRC-8. The host may stop after any word with a NACK and a STOP. The driver checks every CRC and returns DriverError::Crc on the first mismatch.
  • The CRC-8 covers the two bytes before it: polynomial 0x31 (x⁸ + x⁵ + x⁴ + 1), initial value 0xFF, no reflection, no final XOR. Test vector: CRC(0xBEEF) = 0x92, checked at compile time by a static_assert.

Commands

Code Command (argument) Driver call, notes
0x3603 Start, O2 StartContinuous(Gas::O2)
0x3608 Start, air StartContinuous(Gas::Air)
0x3615 Start, N2O StartContinuous(Gas::N2O); SFM4300-20 only
0x361E Start, CO2 StartContinuous(Gas::CO2); SFM4300-20 only
0x3632 Start, air / O2 (O2 ‰) StartContinuous(Gas::AirO2Mix, o2)
0x3639 Start, N2O / O2 (O2 ‰) StartContinuous(Gas::N2OO2Mix, o2); SFM4300-20 only
0x3646 Start, CO2 / O2 (O2 ‰) StartContinuous(Gas::CO2O2Mix, o2); SFM4300-20 only
0xE17D Transmit concentration (O2 ‰) UpdateConcentration(); while measuring
0xE000 Reset I2C address pointer UpdateConcentration(), after 0xE17D
0x3FF9 Stop continuous measurement Stop(); idle after up to 0.5 ms, which Stop() waits
0x366A Configure averaging (N) ConfigureAveraging(); idle only; 0 = average-until-read, 1 to 128 = fixed N
0x3661 Read scale factor, offset, unit (start code) ReadScaleOffsetUnit(); idle only; three words back
0x3677 Enter sleep EnterSleep(); idle only
(address + W) Exit sleep ExitSleep(); not acknowledged, poll until the address ACKs (typically 16 ms)
0xE102 Read product identifier ReadProductIdentifier(); idle only; six words back
0x06 to 0x00 Soft reset (general call) SoftReset(); about 16 ms, 20 ms maximum

Codes 0x3624 and 0x362F are reserved on the SFM4300. While a continuous measurement runs, the sensor accepts only stop, soft reset and the two concentration commands: stop it before anything else.

Frames the sensor returns

After Words (each + CRC) Bytes
A start command flow, temperature, status 9 (3 for flow only)
0x3661 scale factor, offset, unit 9
0xE102 product number (2 words), serial number (4 words) 18

Flow, temperature, scale factor and offset are signed 16-bit (two's complement); the unit and the status are unsigned. A read header while no new sample exists is NACKed (DriverError::NoData).

Scaling

The sensor reports raw counts. The physical value is (raw - offset) / scale, with the scale factor and offset of the gas table in use:

Signal Scale factor Offset Unit
Flow, SFM4300-20 (every gas and mixture) 2500 -28672 slm
Flow, SFM4300-50 (O2, air, air/O2) 1000 -28672 slm
Temperature 200 0 °C

The driver does not hard-code the flow scaling: StartContinuous() reads it from the sensor (0x3661) for the gas you start, and ReadMeasurement() applies it. The temperature scale (200 per °C) is a constant, sfm::kTemperatureScalePerC.

The flow unit is 0x0148 on every SFM4300 table: standard litres per minute at 20 °C and 1013.25 mbar (bits 12:8 = 1 standard litre, bits 7:4 = 4 per minute, bits 3:0 = 8 prefix 10⁰).

Status word

The third word of each measurement frame.

Bits Name Meaning
15:12 running command Index of the start command in the table below
11 exponential smoothing 1: average-until-read has switched to exponential smoothing (no read for over 64 ms)
10 averaging mode 0: average-until-read; 1: fixed-N averaging
9:0 gas concentration O2 volume fraction in ‰ for a mixture; 0x3FF for a pure gas
Bits 15:12 Start code Measurement
0 0x3603 O2
1 0x3608 Air
2 0x3615 N2O
3 0x361E CO2
4, 5 0x3624, 0x362F Reserved
6 0x3632 Air / O2
7 0x3639 N2O / O2
8 0x3646 CO2 / O2

sfm::StatusWord decodes the fields: command_nibble(), exponential_smoothing(), fixed_n_averaging(), gas_fraction_permille() and pure_gas().

sfm::GasFromStatusNibble(family, nibble) turns bits 15:12 back into an sfm::Gas using this table (SF06) or the SFx6000 one, and sfm::StatusNibbleIsKnown(family, nibble) tells a real code from a reserved one: a reserved nibble decodes as Gas::Air. On the SFM3013-300-CLM, 2 and 7 are HeOx and HeOx / O2: the GasFromStatusNibble(variant, nibble) overload decodes them.

Product identifier

Bytes 1 to 4 of the 0xE102 reply form a 32-bit product number; the last 8 bits are a revision. The driver ignores them when it decodes the variant.

Product number Variant sfm::Variant
0x0403011x SFM4300-20-B Sfm4300_20_B
0x0403021x SFM4300-20-O Sfm4300_20_O
0x0403031x SFM4300-20-P Sfm4300_20_P
0x0403091x SFM4300-50-B Sfm4300_50_B
0x0403071x SFM4300-50-O Sfm4300_50_O
0x0403061x SFM4300-50-P Sfm4300_50_P

The datasheet marks ...8x revisions as prototypes and ...1x as production parts. The 64-bit serial number, read as a decimal, has the form yywwxxxxxx: calibration year and week, then a sequence number.