The settings a BASIS-2 takes over Modbus, their ranges and defaults, and the order to apply them in.

Every setting here is a register write that the instrument keeps until you change it or restore the factory settings. Every snippet is compiled from tests/docs/configuration.cpp.

Order

Step Call Why this position
1 SetModbusAddress(), SetBaudRate() Bus settings first, one instrument at a time (API guide)
2 ReadIdentity() Flow units and full scale decide how to read every value
3 SetGas() Calibration for the process gas
4 SetReferenceTemperatureC(), SetFlowAveragingMs(), SetTotalizerLimitMode() How flow and total are reported
5 SetSetpointSource(), SetCommWatchdogMs(), SetAutotareEnabled(), SetMaxSetpointRamp() Controllers: control behaviour
6 Tare() Last, on the process gas with no flow

Meters and controllers

using al::DriverError;
DriverError e = meter.SetGas(al::Gas::Nitrogen).error;                     // register 2100
if (e == DriverError::None) e = meter.SetReferenceTemperatureC(20.0f).error;  // 52: 0..30 C
if (e == DriverError::None) e = meter.SetFlowAveragingMs(50).error;         // 55: 0..2500 ms
if (e == DriverError::None) {
  e = meter.SetTotalizerLimitMode(al::TotalizerLimitMode::ResetWithOvr).error;  // 54
}
return e == DriverError::None;
Setting Call Register Range Factory default
Gas SetGas(Gas) 2100 Air, Argon, CarbonDioxide, Nitrogen, Oxygen, NitrousOxide, Hydrogen, Helium, Methane Per order
Reference temperature SetReferenceTemperatureC(float) 52 0 °C to 30 °C, in 0.01 °C Not stated
Flow averaging SetFlowAveragingMs(uint16_t) 55 0 ms to 2500 ms time constant Not stated
Totalizer limit SetTotalizerLimitMode(TotalizerLimitMode) 54 See below Not stated
ASCII unit ID SetAsciiUnitId(char) 46 'A' to 'Z' 'A'

The driver checks the ranges of the averaging time, the reference temperature, the address and the unit ID before it sends anything, and returns InvalidParameter or InvalidAddress. Writing a gas the instrument does not have leaves the gas unchanged (manual, register 2100).

TotalizerLimitMode At the maximum total
SaturateNoError (0) Hold the maximum, no OVR
ResetNoError (1) Restart from zero, no OVR
SaturateWithOvr (2) Hold the maximum, set OVR
ResetWithOvr (3) Restart from zero, set OVR

Controllers only

using al::DriverError;
DriverError e = controller.SetSetpointSource(al::SetpointSource::DigitalUnsaved).error;  // 516
if (e == DriverError::None) e = controller.SetCommWatchdogMs(500).error;    // 514: 0..5000 ms
if (e == DriverError::None) e = controller.SetAutotareEnabled(true).error;  // 515
if (e == DriverError::None) {
  // 524-525: % of full scale per ms x 10^7. 0.01 %/ms (full scale in 10 s) = 100000.
  e = controller.SetMaxSetpointRamp(100000).error;
}
return e == DriverError::None;
Setting Call Register Range
Setpoint source SetSetpointSource(SetpointSource) 516 Analog (0), DigitalSaved (1), DigitalUnsaved (2)
Communication watchdog SetCommWatchdogMs(uint16_t) 514 0 ms (off) to 5000 ms
Autotare SetAutotareEnabled(bool) 515 Tares after the setpoint has been 0 for at least 2 s
Maximum setpoint ramp SetMaxSetpointRamp(uint32_t) 524-525 % of full scale per ms × 107; 0 = no ramp
Batch volume SetTotalizerBatch(uint32_t) 521-522 Total-volume units, scaled by the factory-set decimals

A meter has none of these registers and rejects the writes (API guide).

Driver settings

These live in the Driver object, not in the instrument.

Setting Call Default
Target address constructor, SetAddress() As constructed
Response timeout constructor, SetTimeoutMs() 200 ms (kDefaultTimeoutMs)
Bus baud rate, for the inter-frame gap SetBusBaudRate(); SetBaudRate() sets it too 38400 (kDefaultBusBaudBps)
Flow decimals argument of ReadInstantaneous() 3

Back to factory settings

FactoryRestore() writes 0x5214 to register 80 and copies the factory settings over the user settings. The manual recommends a power cycle afterwards. It may change the address and the baud rate back, so expect to rediscover the instrument.