Alicat BASIS-2 Driver 1.0.0
API reference
|
Main BASIS-2 driver. More...
#include <alicat_basis2.hpp>
Public Member Functions | |
| Driver (UartT &uart, uint8_t address, uint16_t timeout_ms=kDefaultTimeoutMs) noexcept | |
| Construct a driver bound to one bus + one slave address. | |
| Driver (const Driver &)=delete | |
| Driver & | operator= (const Driver &)=delete |
| void | SetTimeoutMs (uint16_t ms) noexcept |
| Update the per-transaction timeout (ms). | |
| uint16_t | GetTimeoutMs () const noexcept |
| void | SetBusBaudRate (uint32_t bps) noexcept |
| Tell the driver the bus baud rate (default 38400, the factory rate). | |
| uint32_t | GetBusBaudRate () const noexcept |
| uint32_t | GetInterFrameGapUs () const noexcept |
| void | SetAddress (uint8_t addr) noexcept |
| Switch the slave address this driver targets at runtime. | |
| uint8_t | GetAddress () const noexcept |
| DriverResult< InstantaneousData > | ReadInstantaneous (uint8_t flow_decimals=3) noexcept |
| Read all instantaneous registers (2100..2109) in one burst. | |
| DriverResult< InstrumentIdentity > | ReadIdentity () noexcept |
| Pull the static identification of the instrument. | |
| DriverResult< void > | SetSetpoint (float value_in_user_units, uint8_t flow_decimals=3) noexcept |
| Set the digital setpoint (controllers only). | |
| DriverResult< void > | SetSetpointSource (SetpointSource src) noexcept |
| Switch the setpoint source (controllers only). | |
| DriverResult< void > | SetCommWatchdogMs (uint16_t ms) noexcept |
| Configure the communication watchdog (controllers only). | |
| DriverResult< void > | SetMaxSetpointRamp (uint32_t pct_per_ms_x_10e7) noexcept |
| DriverResult< void > | Tare () noexcept |
| Force a tare (zero-flow reference) on the instrument. | |
| DriverResult< void > | SetAutotareEnabled (bool enabled) noexcept |
| Enable / disable automatic taring (controllers only). | |
| DriverResult< void > | SetGas (Gas g) noexcept |
| Select one of the 9 factory-calibrated gases. | |
| DriverResult< void > | ResetTotalizer () noexcept |
| Reset the totalizer to zero. | |
| DriverResult< void > | SetTotalizerLimitMode (TotalizerLimitMode mode) noexcept |
| Configure totalizer overflow behaviour. | |
| DriverResult< void > | SetTotalizerBatch (uint32_t value_scaled) noexcept |
| Set the totalizer batch volume (controllers only). | |
| DriverResult< void > | SetFlowAveragingMs (uint16_t ms) noexcept |
| Configure on-chip flow averaging (IIR-style time constant). | |
| DriverResult< void > | SetReferenceTemperatureC (float c) noexcept |
| Set the reference temperature used for standard / normal flow. | |
| DriverResult< void > | ConfigureMeasurementTrigger (uint16_t trigger_bits) noexcept |
| Configure when a new measurement window auto-restarts. | |
| DriverResult< void > | StartMeasurementSamples (uint16_t num_samples) noexcept |
Start a measurement window of num_samples samples. | |
| DriverResult< MeasurementData > | ReadMeasurement (uint8_t flow_decimals=3) noexcept |
| Read the current / most-recent measurement window result. | |
| DriverResult< void > | SetModbusAddress (uint8_t new_address) noexcept |
| Change the instrument's Modbus slave address. | |
| DriverResult< void > | SetAsciiUnitId (char id) noexcept |
| Change the instrument's ASCII unit-id (used by the legacy ASCII serial protocol). | |
| DriverResult< void > | SetBaudRate (BaudRate br) noexcept |
| Change the serial baud rate (instrument register 21). | |
| DriverResult< void > | FactoryRestore () noexcept |
| Restore the instrument to factory defaults. | |
| DriverResult< uint8_t > | DiscoverPresentAddresses (uint8_t *present_bitmap, std::size_t bitmap_bytes, uint16_t probe_timeout_ms=30, uint8_t first_addr=1, uint8_t last_addr=kMaxModbusAddress) noexcept |
| Probe the entire bus to find which addresses respond. | |
| DriverError | ReadHoldingRegisters (uint16_t reg_addr, uint16_t *out, uint8_t count) noexcept |
Function 03: read count holding registers starting at reg_addr. | |
| DriverResult< void > | WriteSingleRegister (uint16_t reg_addr, uint16_t value) noexcept |
| Function 06: write one holding register. | |
| DriverResult< void > | WriteMultipleRegisters (uint16_t reg_addr, const uint16_t *regs, uint8_t count) noexcept |
Function 16: write count (1..32) consecutive holding registers. | |
| ModbusExceptionCode | GetLastException () const noexcept |
| Exception code of the last transaction's exception reply. | |
Main BASIS-2 driver.
| UartT | Concrete CRTP adapter inheriting UartInterface<UartT>. |
The driver instance corresponds to one Modbus address on the shared RS-485 bus. Multiple driver instances can share the same UART adapter for different addresses; the host is responsible for serialising bus access (the hf-core handler wraps this in an RtosMutex).
|
inlinenoexcept |
Construct a driver bound to one bus + one slave address.
| uart | Reference to a CRTP adapter; must outlive Driver. |
| address | Modbus address (1..247) of this slave. Use kBroadcastAddress (0) for unaddressed broadcast writes (response-less). |
| timeout_ms | Per-transaction response timeout (datasheet default ~200 ms is safe). |
|
delete |
|
inlinenoexcept |
Configure when a new measurement window auto-restarts.
| trigger_bits | OR-combination of MeasurementTrigger::* flags (OnDigitalSetpointChange, OnHoldPercentChange, OnAverageFlowRead). 0 disables auto-trigger. |
|
inlinenoexcept |
Probe the entire bus to find which addresses respond.
| [out] | present | Bitmap; bit N → instrument at address N. |
| probe_timeout_ms | Per-address timeout (defaults to short). | |
| first_addr | / last_addr Sweep window (default full 1..247). |
Issues a Function-03 read of register 25 (firmware version) to every address in [first_addr, last_addr]. Restores this driver's saved address on exit.
|
inlinenoexcept |
Restore the instrument to factory defaults.
Writes the magic value 0x5214 to register 80. The datasheet recommends a power-cycle afterwards to make sure every cached value picks up the reset.
|
inlinenoexcept |
|
inlinenoexcept |
|
inlinenoexcept |
|
inlinenoexcept |
Exception code of the last transaction's exception reply.
DriverError::ModbusException), ModbusExceptionCode::None otherwise. Reset at the start of every request.
|
inlinenoexcept |
|
delete |
|
inlinenoexcept |
Function 03: read count holding registers starting at reg_addr.
None, or the transaction's error. An exception reply returns ModbusException as soon as its 5 bytes arrived; its code is in GetLastException().
|
inlinenoexcept |
Pull the static identification of the instrument.
Reads firmware version (reg 25), serial number (regs 26..31), Modbus address (reg 45), ASCII unit id (reg 46), full-scale flow in user units (regs 47..48) and the flow-units code (reg 49).
DriverResult<InstrumentIdentity> populated with all of the above on success; transport / Modbus error otherwise.
|
inlinenoexcept |
Read all instantaneous registers (2100..2109) in one burst.
This is the canonical "live data" call — one round-trip on the bus, decoded into engineering units using the flow_decimals factor read once from the instrument.
| flow_decimals | The factory-set number of decimals reported on this instrument's flow register; pulled at init time via ReadIdentity(). |
DriverResult<InstantaneousData> — value carries the decoded snapshot on success, error is DriverError::None. On failure value is default-constructed and error carries the underlying transport / Modbus error.
|
inlinenoexcept |
Read the current / most-recent measurement window result.
One request reads registers 4201..4215 (manual Rev 4, "Measurement Data Registers", pp. 22-23). The previous window's average flow is the signed 32-bit value at 4214-4215, high register first.
| flow_decimals | The instrument's factory-set flow decimals, used for avg_flow (as in ReadInstantaneous()). |
DriverResult<MeasurementData> populated with min / max / average temperature and flow on success; transport / Modbus error on failure. MeasurementTrigger::OnAverageFlowRead set, reading the low register 4215 starts the next measurement window.
|
inlinenoexcept |
Reset the totalizer to zero.
Writes 0xAA55 to register 53. Total volume since the last power-up / reset is cleared and any active totalizer-overrange (OVR) bit is also dropped.
|
inlinenoexcept |
Switch the slave address this driver targets at runtime.
|
inlinenoexcept |
Change the instrument's ASCII unit-id (used by the legacy ASCII serial protocol).
| id | A capital letter ‘'A’..'Z'. @return Success or transport / Modbus error; DriverError::InvalidParameterifidis outsideA..Z`. |
|
inlinenoexcept |
Enable / disable automatic taring (controllers only).
When enabled, the instrument tares automatically after the setpoint has been at 0 for at least two seconds.
| enabled | true to enable autotare, false to disable. |
|
inlinenoexcept |
Change the serial baud rate (instrument register 21).
After the write the driver uses the new rate for its inter-frame gap (SetBusBaudRate()), whatever the result: the instrument switches as soon as it receives the request.
|
inlinenoexcept |
Tell the driver the bus baud rate (default 38400, the factory rate).
Only used for the Modbus-RTU inter-frame gap: before every request the driver waits modbus::InterFrameGapMs(bps) through the port's delay_ms() (3.5 characters, 1.75 ms above 19200 baud; Modbus over Serial Line V1.02 section 2.5.1.1), so the request cannot start inside the idle gap that ends the previous frame. SetBaudRate() updates it. Several drivers on one port: set it on each.
|
inlinenoexcept |
Configure the communication watchdog (controllers only).
| ms | 0..5000 ms; 0 disables. The instrument forces a zero setpoint and closes the valve if no Modbus traffic is seen within this window when the setpoint source is DigitalUnsaved (see datasheet "Communication Watchdog"). |
DriverError::InvalidParameter if ms > 5000.
|
inlinenoexcept |
Configure on-chip flow averaging (IIR-style time constant).
| ms | Time constant in milliseconds, 0..2500. Higher values smooth fluctuations at the cost of response. |
DriverError::InvalidParameter if ms > 2500.
|
inlinenoexcept |
Select one of the 9 factory-calibrated gases.
| g | One of Gas::Air / Argon / CarbonDioxide / Nitrogen / Oxygen / NitrousOxide / Hydrogen / Helium / Methane. |
|
inlinenoexcept |
|
inlinenoexcept |
Change the instrument's Modbus slave address.
| new_address | New slave address, 1..247 (kMaxModbusAddress). |
DriverError::InvalidAddress if new_address is outside the legal range.
|
inlinenoexcept |
Set the reference temperature used for standard / normal flow.
| c | Temperature in °C, 0..30. |
DriverError::InvalidParameter if out of range.
|
inlinenoexcept |
Set the digital setpoint (controllers only).
Writes the signed 32-bit setpoint to registers 2053..2054. The datasheet defines the on-the-wire scaling as value × 1000 so the host passes the engineering value (e.g. 12.5 SLPM) and the driver does the multiplication.
| value_in_user_units | Real setpoint, e.g. 12.5 SLPM. |
| flow_decimals | Number of decimals for this instrument (call ReadIdentity once and cache). |
DriverResult<void> — success or a Modbus / transport error. A meter (non-controller) instrument will return DriverError::ModbusException.
|
inlinenoexcept |
Switch the setpoint source (controllers only).
| src | Analog input pin / digital saved-to-flash / digital volatile. |
|
inlinenoexcept |
Update the per-transaction timeout (ms).
|
inlinenoexcept |
Set the totalizer batch volume (controllers only).
Once the totalizer reaches the batch volume the controller closes the valve. Pass the value already pre-scaled per the instrument's factory-set decimal count.
| value_scaled | Batch volume × 10^decimals (per datasheet). |
|
inlinenoexcept |
Configure totalizer overflow behaviour.
| mode | SaturateNoError / ResetNoError / SaturateWithOvr / ResetWithOvr. |
|
inlinenoexcept |
Start a measurement window of num_samples samples.
Each sample is taken at approximately a 2.5 ms interval. Writing the register cancels any in-progress window and shifts the accumulated data into the "previous measurement" slot.
| num_samples | Number of samples to collect (unsigned 16-bit). |
|
inlinenoexcept |
Force a tare (zero-flow reference) on the instrument.
Writes the magic value 0xAA55 to register 39. The instrument must be exposed to true zero flow with the actual process gas in the line for the result to be in-spec.
|
inlinenoexcept |
Function 16: write count (1..32) consecutive holding registers.
ModbusException as soon as its 5 bytes arrived; its code is in GetLastException().
|
inlinenoexcept |
Function 06: write one holding register.
ModbusException as soon as its 5 bytes arrived; its code is in GetLastException().